PURPOSE: Automated seizure detection at home is mostly done using either patient-independent algorithms or manually personalized algorithms. Patient-independent algorithms, however, lead to too many false alarms, whereas the manually personalized algorithms typically require manual input from an experienced clinician for each patient, which is a costly and unscalable procedure and it can only be applied when the patient had a sufficient amount of seizures. We therefore propose a nocturnal heart rate based seizure detection algorithm that automatically adapts to the patient without requiring seizure labels.

METHODS: The proposed method initially starts with a patient-independent algorithm. After a very short initialization period, the algorithm adapts to the patients’ characteristics by using a low-complex novelty detection classifier. The algorithm is evaluated on 28 pediatric patients with 107 convulsive and clinical subtle seizures during 695?h of nocturnal multicenter data in a retrospective study that mimics a real-time analysis.

RESULTS: By using the adaptive seizure detection algorithm, the overall performance was 77.6% sensitivity with on average 2.56 false alarms per night. This is 57% fewer false alarms than a patient-independent algorithm with a similar sensitivity. Patients with tonic-clonic seizures showed a 96% sensitivity with on average 1.84 false alarms per night.

CONCLUSION: The proposed method shows a strongly improved detection performance over patient-independent performance, without requiring manual adaptation by a clinician. Due to the low-complexity of the algorithm, it can be easily implemented on wearables as part of a (multimodal) seizure alarm system.

Related News